參考價格
面議型號
品牌
產(chǎn)地
江蘇樣本
暫無看了ASTM E162熱輻射火焰?zhèn)鞑y試儀的用戶又看了
虛擬號將在 180 秒后失效
使用微信掃碼撥號
儀器簡介:
The Radiant Panel Flame Spread Tester is an equipment using gas ignited Radiant heat panel to test the surface combustibility of Building product (ASTM E 162) and cellular plastics (ASTM E 365), Radiant Panel Index(Is) is measured by the flame radio waves, Critical Heat Flux,(CHF) and Heat Release Rate(kW/m2).
技術(shù)參數(shù):
• ASTM E 162: Standard Test Method for surface Flammability of materials Using a
Radiant Heat Energy Source
• ASTM D 3675: Standard Test Method for surface Flammability of Materials using
A Radiant Heat Energy Source
主要特點(diǎn):
• Reflection panel is composed of a cast iron frame with vertically applied
porosity fire resistant material with a 305X457mm size exposed reflection surface
and is designed to function at 815 ℃.
• The Calibration burner should have 1.8mm holes in 16mm intervals, and should
be manufactured along the middle line.
• An exhaustion device with a 0.5 m/s airflow exhaustion rate is installed.
• In order to control the surface heat of the reflection panel, a 180℃ ~ 230℃
(Black body temperature) range will be given, and a ±5℃ accuracy, fit to observe
a surface of 250mm radius at 2 m distance.
• Base Structure Units
- Frame dimension : 1850 (H) X 700 (D) X 1250 (L)
- Electrical Power : 220 V AC, 60 Hz, 12 A
- Sliding platform for loading of the test specimen
- Radiant Panel (90°, Porous cement)
- 19" Touch Screen Panel Computer for control & Acquisition Data
• Radiant Panel with Air and Gas Supply
- Radiant Panel Size : 300 by 460 (mm)
- Fine Brick : Max Temp : 815℃ (1500℉)
- Pilot Burner : Acetylene(0.5 l/min, 150 mbar)
- Calibration Burner : Methane gas(95% purity, 20 l/min, 150 mbar)
- Propane Burner : Pressure(1.0 l/s, 20 mbar)
- Compressed Air Burner : 50 l/s, 700 Pa of water
- Digital anemometer : 0.5 m/s
• Analysis & Controls
- Heat Flux Meter : Schmidt - Bolter type, Range : 0 ~ 100 ± 10 ?/㎡
- Portable water cooling system for heat flux meter
- Pyrometer for measuring radiant panel temperature
- DAQ (Data Acquisition) System(GPIB/USB Interface)
• Application Software
- ASTM E 162 Operation & DAQ Program
- Test Calibration Program
暫無數(shù)據(jù)!